Thermometer Calibration - Infrared and Handheld
Infrared thermometer calibration of instruments such as the Fluke, Meco and Thermal Imager Liquid in Glass and Digital/Handheld Thermometers of all types ranges. Field/on-site measurements availableAction Engineers is a NABL accredited thermal & humiditycalibration lab. We offer one of the most comprehensive thermal and infraredcalibration across India. Our ISO 17025 calibration scope allows us tocalibrate Temperature Indicator / Controller / Data Logger / Scanner /Transmitters, Dial Temperature Gauge, Glass Thermometer, IR Thermometers andmore.
Our scope also allows us to calibrate Humidity Chamber,Humidity Indicator / Controller / Data Logger / Scanner / Transmitters,Humidity – Validation / Mapping Auto Clave and much more.
Our digitally enabled unique features:
On-site calibration is also offered by Action Engineers. We make sure the calibration services we provide at your location follow the same quality standards as followed by us at our calibration labs.
For more information, Request a Quote today!
| S.No | Discipline / Group | Measurand or Reference Material / Type of instrument or material to be calibrated or measured / Quantity Measured / Instrument | Calibration or Measurement Method or procedure | Measurement range and additional parameters where applicable (Range and Frequency) | Calibration and Measurement Capability(CMC) (±) |
| 1 | THERMAL-SPECIFIC HEAT & HUMIDITY | Environmental Chamber @ 50 %RH | Using Humidity/Temperature Data Logger with Sensor (9 Numbers) By spatial Mapping Method as per DKDR5-7 | 10 °C to 50 °C | 2.44°C |
| 2 | THERMAL-SPECIFIC HEAT & HUMIDITY | Humidity Chamber / Environmental Chamber @25°C | Using Humidity/Temperature Data Logger with Sensor (9 Numbers) By Spatial Mapping Method as per DKDR5-7 | 30 %RH to 90 %RH | 4.2%RH |
| 3 | THERMALSPECIFIC HEAT & HUMIDITY | Humidity Indicator of Humidity Chamber / Environmental Chamber @25°C (Single Position) | Using Humidity/Temperature Indicator with Sensor -By Direct Method | 30 % RH to 90%RH | 3.09%RH |
| 4 | THERMALSPECIFIC HEAT & HUMIDITY | Thermo-Hygrometer, Humidity Indicator / Controller / Transmitter / Data Logger / Recorder with Sensor @25°C | Using Humidity Indicator with Sensor & Temperature - Humidity Chamber - By Comparison Method | 30%RH to 90% RH | 3.09%RH |
| 5 | THERMALSPECIFIC HEAT & HUMIDITY | Thermo-Hygrometer, Temperature Humidity Indicator /Controller / Transmitter / Scanner / Recorder / Data Logger with Sensor @ 50 %RH | Using Temperature Indicator with Sensor, 6.5 digits DMM temperaturehumidity Chamber By Comparison Method | 10 °C to 50 °C | 0.82°C |
| 6 | THERMAL-TEMPERATURE | Digital, / Analogue thermometer, Thermocouple, RTD Sensor,Temperature Gauge, Temperature Sensor with Indicator /Controller / Scanner/ Transmitter / Data Logger / Thermal Switch | Using RTD-Sensor With Indicator & Liquid Bath , 51/2 Multimeter- By Comparison Method | -35 °C to 50 °C | 0.27°C |
| 7 | THERMAL TEMPERATURE | Freezer / Cold Temperature Chamber | Using RTD Sensors (9 numbers) With Temperature Scanner By Spatial Mapping Method as per DKD R5-7 | -35 °C to 50 °C | 2 °C |
| 8 | THERMAL TEMPERATURE | Hot Air Oven / Furnace / Temperature Chamber /(Autoclave- for non medical purpose only) | Using RTD Sensors (9 numbers) With Temperature Scanner By Spatial Mapping Method as per DKD - R5-7 | 30 °C to 300°C | 2.0°C |
| 9 | THERMAL TEMPERATURE | Infrared Thermometer / Pyrometer / Thermal Image Camera @ emissivity 0.95 | Using IR Thermometer and Black Body Source with Emissivity 0.95 - By Comparison Method | 50 °C to 500 °C | 1.8°C to 3.4°C |
| 10 | THERMAL TEMPERATURE | Liquid Bath / Water Bath / (Incubator - for non medicalpurpose only) | Using RTD Sensors (9 numbers) With Temperature Scanner By Spatial Mapping Method as per DKD - R5-7 | -35 °C to 250 °C | 2 °C |
| 11 | THERMAL TEMPERATURE | Liquid In Glass Thermometer / Dial Temperature Gauge | Using RTD Sensor With Indicator & Temperature Liquid Bath By Comparison Method | -35 °C to 250 °C | 0.84 °C |
| 12 | THERMAL TEMPERATURE | Temperature Indicator of Freezer / Temperature Indicator of Rotary Shaker & Centrifuge (Single Position) | Using RTD- Sensor with Indicator By Direct Method | -80 °C to 50 °C | 0.5 °C |
| 13 | THERMAL TEMPERATURE | Temperature Indicator of Liquid Bath / Water Bath / (Incubator- for non medical purpose only) (Single Position) | Using RTD- Sensor with Indicator By Direct Method | -35 °C to 250 °C | 0.58°C |
| 14 | THERMAL TEMPERATURE | Temperature Indicator of Oven / Dry Block Temperature Calibrator / Furnace / Temperature Chamber (Single Position) | Using RThermocouple, with Temperature Indicator By Direct Method | 400 °C to 1200 °C | 1.77°C |
| 15 | THERMAL TEMPERATURE | Temperature Indicator of Oven/Dry Block TemperatureCalibrator/Furnace/Temperature Chamber / Temperature Indicator of Rotary Shaker & Centrifuge/ (Autoclave / Incubator - for non medical purpose only | Using RTD- Sensor with Indicator By Direct Method | 30 °C to 400 °C | 0.5 °C |
| 16 | THERMAL TEMPERATURE | Thermometer, Thermocouple, RTD Sensor, Temperature Gauge, Temperature Sensor with Indicator / Controller / Scanner / Transmitter / Data Logger / Thermal Switch | Using Thermocouple / RTD-Sensor With Indicator Dry Block Calibrator, 51/2 Multimeter - By Comparison Method | 50 °C to 400 °C | 0.6°C |
| 17 | THERMAL TEMPERATURE | Thermometer, Thermocouple, Temperature Gauge, Temperature Sensor with Indicator / Controller / Scanner / Transmitter / Data Logger / Thermal Switch | Using Thermocouple With Indicator & Dry Block Calibrator, 51/2 Multimeter - By Comparison Method | 400 °C to 1.6 °C | 1.6°C |
| 18 | THERMAL TEMPERATURE | Thermometer, Thermocouple, Temperature Gauge, Temperature Sensor with Indicator / Controller / Scanner/ Transmitter / Data Logger / Thermal Switch | Using RThermocouple With Indicator & Dry Block Calibrator , 5 1/2 Multimeter- By Comparison Method | 600 °C to 1200 °C | 1.8°C |